Blue Dragon Model C680 Combustion TOC Analyzer
Catalytic combustion and NDIR for TOC, TC, IC and NPOC in one bench instrument.
A laboratory analyzer using high-temperature catalytic combustion with NDIR detection to measure TOC, TC, IC and NPOC across water matrices from purified water through to industrial wastewater.
ACTS is Blue Dragon Technology’s sole agent in Egypt, supplying the Model C680 High-Temperature Combustion TOC Analyzer to oil & gas, petrochemical and power generation plants nationwide from our office in Sheikh Zayed City, Giza. Sizing and selection are handled by our own application engineers, with aftermarket support and technical services once the unit is in service.
- Furnace 680 to 1000 °C with catalytic combustion
- NDIR CO₂ detection
- Manual microsyringe injection, 10 to 500 µL
- TOC, TC, IC and NPOC from one instrument
| Detection Limit | ≤0.5 mg/L (HJ 501-2009) |
|---|---|
| Repeatability | ≤3% |
| Analysis Time | 4 to 10 min |
| Range | 0 to 100,000 mg/L (method dependent) |
| TC / IC Error | ±0.1% FS or ±5% / ±0.1% FS or ±4% |
| Linearity | R² ≥99.9% |
| Ambient | 10 to 40 °C, 10 to 85% RH |
| Power | 220–240 VAC 50 Hz, ≈1200 W heating |
| Utilities | Carrier gas O₂ ≥99.99%, reagent 0.1 mol/L H₃PO₄ |
Specifications transcribed from the manufacturer datasheet. Confirm final selection with the product catalog or on request.
Model C680 High Temperature Combustion TOC Analyzer Product Sheet
Model specification sheet with ranges, outputs and utilities.
Model C680 High Temperature Combustion TOC Analyzer Brochure
Product overview and selection guide.
Need a Model for your project?
Sized, quoted, and answered by an engineer
Send us your requirement and one of our application engineers will follow up, typically within 24 hours.